Temperature monitoring
Corrective action
Possible causes
Problem
1
Switch to the second input of the
temperature probe if possible.
2
If the second input is not present
or is not working, replace the
probe.
Temperature probe has failed.
Temperature monitor-
ing is not working as
expected.
Insert temperature probe into the
thermowell.
Temperature probe is not inserted
into the disposable bag.
Weight monitoring
Corrective action
Possible causes
Problem
Connect the load cell.
Load cell is disconnected.
Load cell is not functioning as
expected.
Contact the metrology depart-
ment to perform a cornering test
and determine which load cell
has failed.
Load cell has failed (physically
or electrically broken).
Verify that the mixer vessel legs
are free of any obstructions or
blocking objects.
Physical obstruction of load cell
motion.
Disengage the jacking bolt. See
Disengage jacking bolts, on
page 135 for instructions.
Note:
If the mixer was filled while the
jacking bolt was engaged, the
weight measurements of the
batch are not accurate.
Jacking bolt is engaged.
Contact your GE representative.
The configuration on the load
cell transmitter is incorrect or
corrupt.
Schedule the calibration of the
load cells.
Load cells require calibration.
Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD
271
9 Troubleshooting