Action
Step
Set the circulating bath temperature to one of the calibration end point
temperatures; or use another bath if available. Repeat steps 3 to 5 above.
6
Set the circulating bath temperature to the other calibration end point
temperature; or use another bath if available. Repeat steps 3 to 5 above.
7
Evaluate the obtained results considering the following:
8
•
Manufacturer's tolerance: ± 0.5°C
•
Process tolerance: determined by the end user.
Adjustment
If the temperature probe is out of tolerance, follow the steps below to adjust the probe.
Action
Step
Using the data obtained during the calibration procedure, calculate the
offset to adjust the probe temperature to the recorded standard probe
temperature:
1
Offset = standard temperature reading - probe reading.
Note:
Use the data obtained at operating temperature (the middle calibration point)
for best overall accuracy.
If the probe contains two temperature sensors, calculate the offsets for both
sensors.
2
Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD
291
10 Reference information
10.8 Temperature probe calibration