2.3.1
Emergency procedures, Xcellerex XDM mixer
Emergency shutdown of XDM
mixer
Result
Action
The agitator motor is immediately stopped. All
power is removed from the mixer.
Switch off the mains power supply by turning the
power switch counterclockwise to the O OFF posi-
tion.
All power is removed from the I/O Cabinet.
If required, disconnect the power cord from the
mains power supply.
Restart XDM mixer
Follow the instructions below to restart XDM mixer after an emergency shutdown or a
power failure.
Action
Step
Make sure that the condition that caused the emergency shutdown is cor-
rected.
1
Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD
35
2 Safety instructions
2.3 Emergency procedures
2.3.1 Emergency procedures, Xcellerex XDM mixer