Instructions for Use: GEV019-100
Page 17
Electrical Connections:
Power supply to the GeVentor is via an external 30 VDC power supply (refer to Figure below). The power supply is a medical grade
power supply with compliance certificates for EMC and RF output testing which comply with AS/NZS 3112. A Portable Appliance
Test (PAT), as per IEC60601 (test and tag) is completed on the GeVentor at the factory. The unit is supplied with an Australia three-
pin plug.
DANGER
Risk of electric shock. GeVentor must be connected
to the earthed main power supply.
WARNING
GeVentor power supply cabling is to be routed in such
a way to avoid tripping hazards.
Battery Back-up:
The GeVentor
is supplied with continuous battery supply effectively “charged” by the power supply
, consisting of two batteries.
The batteries have a typical back-up life of 25 minutes and a minimum charge time of 30 minutes. Refer to Section 6.19 (Battery
Back-up Indicator) for the battery health indicator details.
4.6
Gas Connections
DANGER
Device should only be connected to a pipeline
installation designed using a diversity factor that
allows for the indicated high flow at a specified
number of terminal outlets, to avoid exceeding the
pipeline design flow.
On / Off Switch
Power supply
cable port