Instructions for Use: GEV019-100
Page 43
Should the u
ser elect to switch the GeVentor off (select the “NO” option above), the following
screen appears. The user can select
“YES” or “NO
depending on the user’s requirement.
When it is intended to switch off the ventilator and disconnect all gas supplies:
•
Switch off the ventilator using the ON/OFF switch.
•
Physically check that the GeVentor shuts down.
•
Isolate oxygen and air supply if shutting down for long periods.
DANGER
Ensure that there is no requirement for ventilation
on the patient. When the ventilator is shut down,
no ventilation occurs.
6.18
Unscheduled Stops
After an unscheduled stop, the cause of the stop shall be determined. No restart shall be allowed until the cause is known and
clearance from all stations has been obtained.
NOTE
Manual ventilation may be required while the
machine reboots
–
at the discretion of the treating
healthcare professional.