30
31
Chapter 7 Error Message and Trouble Shooting
Message
Cause
Solution
E-0
There may be a problem
with the meter or code
strip.
Reload the code strip. If the problem
persists, please contact the dealer
for service.
E-9
T he test stri p may b e
removed after applying
blood to the absorbent
hole.
Review the instructions and try again
with a new test strip.
Please don't remove the strip before
the test is completed.
E-A
There is a problem with
the meter.
Review the instructions and try again
with a new test strip.
If the problem persists,please
contact the dealer for service.
E-b
The power of the battery is
too low to run a test.
Replace the battery at once.
Message
Cause
Solution
E-C
Usage of the code strip
was inserted improperly,
damaged, or incorrect.
Check that the code number on the
meter display matches the code
number on the test strip vial. Code
the meter or retest as necessary. If
the error message appears again,
please contact the dealer for service.
E-t &
Tm Sign
T h e t e m p e r a t u r e w a s
below or above the system
operating range.
Repeat the test after the meter
and the test strip have reached the
temperature within the operating
range.
EU
It could be caused by a used
test strip or a temporary
or permanent electronic
problem.
Repeat the test with a new test strip.
If the error message appears again,
please contact the dealer for service.
Summary of Contents for BENECHECK GLU