Chapter 8 Specifications
System Accuracy :
± 20% at glucose level ≥ 75 mg/dL (4.17 mmol/L)
Test Sample :
Capillary whole blood
Measuring Time :
Less than 10 seconds
Measuring Range :
20- 600 mg/dL (1.1 - 33.3 mmol/L)
Sample Volume :
≥ 0.7 µL
Strip Storage Condition :
10 - 30˚C
Operation Temperature :
10 - 40˚C
Relative Humidity :
Less than 95%
Memory :
360 test results
Battery Type :
One 3V (CR2032) lithium battery
Battery Life :
Approximately 1,000 tests
Dimensions :
61*77*19 mm
Weight :
About 47g (batteries not included)
Message
Cause
Solution
Hi
1.Blood glucose level is
higher than 600 mg/dL.
2 .Test strip has problem.
1. Review the instructions and try
again with a new test strip.
2. Check the test strip vial to make sure
that it is not marked beyond the
expiration date.
Lo
1.Blood glucose level is
lower than 20 mg/dL.
2.Test strip has problem.
1. Review the instructions and try
again with a new test strip.
2. Check the test strip vial to make sure
that it is not marked beyond the
expiration date.