Model IR400
15
3.4 Applying Power
Before applying power to the system for the first time, all wiring connections should be checked
for correctness. Upon initial power-up or after a fault condition has been corrected, the unit will
enter a start-up mode for 2 minutes before normal operation. IR400 Ethylene start-up can take
up to 10 minutes. As a unit warms up, output may briefly be negative.
After power is applied, the IR400 should be allowed to stabilize for approximately 60 minutes
while the unit attains the proper operating temperature. After stabilization, it is recommended
that the IR400 be zeroed, per the procedure in Section 4.1 (step 1 only). A gas check should
then be performed to ensure that the unit is operating properly. Use the General Monitors gas
check kit (P/N 32548-x) to perform this check. If the unit does not respond properly, calibrate
per the procedure in Section 4.1.4 (steps 1-4).
•
When connecting the IR400 to a safety system, the +24 VDC (red) wire should be the last
wire connected and first wire disconnected when removing the unit to protect the system
from shorting.
•
If the analog (4-20 mA) output is not used, then the white signal wire must be connected to
ground to prevent a fault condition.
Figure 6a: IR400 with Junction Box
Figure 6b: IR400 with European Junction Box
Summary of Contents for IR400
Page 2: ......
Page 15: ...Model IR400 11 Figure 5b IR400 Mounting Dimensions with a European Box ...
Page 20: ...Model IR400 16 Figure 7 Wiring Diagram from IR400 to Control Equipment ...
Page 41: ...Model IR400 37 9 6 Engineering Documentation ...
Page 45: ...Model IR400 HART Field Device Specification for Point IR Gas Detector ...
Page 78: ...Model IR400 Infrared Point Detector for Hydrocarbon Gas Applications Modbus programming guide ...