Model IR400 HART
17
Response Data Bytes
Byte
Format
Description
0 - 1
Unsigned-16
Mode
2 - 3
Unsigned-16
Sub Mode
4 – 7
Float
AO
8 - 9
Unsigned-16
Priority fault
10 – 11
Bit map
Error status
12
Unsigned-8
Reserved = 0
13
Unsigned-8
Reserved = 0
14
Unsigned-8
Reserved = 0
15
Unsigned-8
Power cycled flag
16
Unsigned-8
Event happened flag
17
Integer-8
% of FS
18-21
Integer-32
Reserved = 0
Command-Specific Response Codes
Code
Class
Description
0
Success
No Command-Specific Errors
1-127
Undefined
Command #164: Get Slow Changing Information
Request Data Bytes
Byte
Format
Description
None N/A
N/A
Response Data Bytes
Byte
Format
Description
0 - 1
signed-16
Temperature
2 - 5
float
Voltage
6 - 7
unsigned-16
% Beam Blockage
Command-Specific Response Codes
Code
Class
Description
0
Success
No Command-Specific Errors
1-127
Undefined