Model TA102A
30
The last
A1
alarm option to appear on the display will be the alarm set point (trip level).
If this level is reached or exceeded the
A1
alarm outputs will activate. The display will
indicate the current
A1
alarm set point (Figure 55). Press the
Mode/Select
switch
repeatedly, until the desired
A1
alarm set point appears on the display (
10
to the
A2
set
point in increments of
5
). The
A1
set point cannot be set higher than the
A2
set point.
30
is the factory default for this selection.
TA102A
0 - 100% LEL
READY
SETUP
SETUP LED
SELECT
MODE/
GENERAL
MONITORS
ALARM SETPOINT
A1
FLASHING A1 LED
FLASHING for SETUP
CHECK MODE,
STEADY for SETUP MODE
Figure 55 – A1 Set Point Option
5.4.6 Fault/Inhibit Option
Next, the user will select the Fault/Inhibit option. The
FAULT
LED on the front panel will
be flashing while the display indicates
Ac
or
nA
(Figure 57). An
Ac
selection specifies
that the Model TA102A will activate the
Fault
circuit while the unit is in the Inhibit Mode.
A
nA
selection specifies that the Model TA102A will not activate its
Fault
circuit when
the unit is placed in the
Inhibit
Mode. A
nA
selection will not disable the
Fault
circuit,
therefore, if a
Fault
occurs during the
Inhibit
Mode, the unit will activate the
Fault
circuit.
Not Active
is the factory default for this selection.