Instruction Manual - Clinic 3.40 Compressors

afm 30 CaRtRIDgE REpLaCEmENt

•  Press the lever down (A), turn of 45° the cup (B) and remove it from the cover (C).
•  Unscrew the filter (D).
•  Screw in the new filter.
•  Insert the cup in the top cover 

and turn 45 °.

af 30 CaRtRIDgE REpLaCEmENt

•  Press the lever down (E), turn of 45° the cup (F) and remove it from the cover (G).
•  Remove the assembled component (H).
•  Unscrew the lower deflector (I).
•  Replace the filter cartridge (L).
•  Screw the lower deflector back on.
•  Insert the assembled 

component in the cup.

•  Insert the cup in the top 

cover top and turn of 45°.

Summary of Contents for CLINIC 3.40 Series

Page 1: ...Instruction Manual Clinic 3 40 Compressors CLINIC 3 40 CLINIC DRY 3 40 H CLINIC DRY 3 40 HS EN...

Page 2: ......

Page 3: ...Dimensions of the machine 18 Technical data 19 Intended use of the machine 20 Source of energy supply 20 Products used for the operation 21 Phases of operation 21 Reasonably foreseeable misuse 21 Tran...

Page 4: ...NFORMATION TO THE CUSTOMER DOCUMENT PAGE TO DO STATEMENT OF DELIVERY OF THE INSTRUCTION MANUAL 5 The authorised operators must complete and sign to confirm the delivery and rea ding of the instruction...

Page 5: ...__ __________ ____________________________________________ Mr _______________________________________ __________ ____________________________________________ Mr _______________________________________...

Page 6: ...______________________________________________________________________________ At the presence of the entrusted operator Mr ____________________________________________________________________________...

Page 7: ...st 2014 30 UE Directive of the European Parliament and of the Council of 26 February 2014 replacing 2004 108 EC Directive of the European Parliament and of the Council of 15 December 2004 on the appro...

Page 8: ...HNICAL DATA The CE marking Fig 1 of the machine indicates compliance with the essential occupatio nal health and safety rules set forth by the European Directives listed in the Declaration of CE Confo...

Page 9: ...er stand all parts of this manual and adhere strictly to the rules herein described in order to assure their own and others safety the best performance of the machine and ensure the maximum efficiency...

Page 10: ...t indicates a requirement obligation to take action Generic prohibition sign It highlights the prohibition to perform an action Danger Sign EX It highlights the risk due to explosion CONSULTATION NOTE...

Page 11: ...nd the turning off of the machine and in total respect of the instructions in this ma nual equipped with personal protective equipment PPD specified and occupying the pla ces see related paragraph OPE...

Page 12: ...tection devices PPD envisioned and occupying the positions described see related paragraph SAFETY REPRESENTATIVE It is a qualified engineer appointed by the customer in possession of technical and pro...

Page 13: ...e par Paragraph Pos Position Rif Reference sec Seconds L Left TAB Table see SEE ALL RIGHTS RESERVED The reserved rights related to this instruction manual are in possession of the manufacturer All inf...

Page 14: ...e www gentilinair com within seven days of purchase and b have completed the registration form completely and accurately filling in the required fields with exact complete and true information and hav...

Page 15: ...n 3 models Tab 1 shows the model names and specifications that distinguish them MODEL DESCRIPTION CLINIC 3 40 Oil free air compressor with 40 litres tank without dryer CLINIC DRY 3 40 H Oil free air c...

Page 16: ...ne KEY 1 Support feet 2 Non return valve 3 Handle 4 Cooling fan protection hood 5 Two cylinder pumping unit 6 Electric motor 7 Restorable circuit breaker 8 Electrovalve 9 Air intake filter 10 Membrane...

Page 17: ...01 07 2019 17 www gentilinair com CLINIC DRY 3 40 H CLINIC DRY 3 40 HS 5 4 5 7 6 3 11 9 1 14 15 16 18 17 2 10 12 12 8 13 5 4 5 7 6 3 11 9 1 14 15 18 17 2 10 12 8 16 Instruction Manual Clinic 3 40 Comp...

Page 18: ...HS A A A B B B C C C Fig 1 Tab 1 Instruction Manual Clinic 3 40 Compressors MODELS A B C clinic 3 40 653 mm 652 mm 455 mm clinic DRY 3 40 H 662 mm 652 mm 455 mm clinic Dry 3 40 HS 662 mm 652 mm 455 mm...

Page 19: ...se db A 67 67 67 Weight Kg 39 40 41 Ambient temperature C 0 40 0 40 0 40 Service class 100 100 100 Motor rotation direction The compressor can be used in both directions of rotation CLINIC 3 40 H CLIN...

Page 20: ...POSITIONING OF THE MACHINE WARNING YOU MUST KEEP THE MACHINE OUT OF REACH OF UNAUTHORIZED PERSONS INTENDED USE Compressed air production to operate dental tools OPERATORS IN CHARGE FOR USE An authori...

Page 21: ...OF USE HAVING THE CHARACTERISTICS INDICATED ON PAR INTENDED USE OF THE MA CHINE IT IS COMPULSORY TO USE THE MACHINE IN THE RESPECT OF THE LEGISLATION IN FORCE IN THE COUNTRY OF USE CONCERNING INTRODUC...

Page 22: ...ANSPORTATION OF THE MACHINE The machine is delivered to the customer through a specialized transportation com pany or by the manufacturer who through its own personnel and appropriate means in complia...

Page 23: ...ABILITY OF LOAD 1 Insert the forks 2 Fig 1 under the pallet 1 keeping them at the centre of the latter 2 Raise the packaging 3 a few centimetres to make sure the load is actually centred in the forks...

Page 24: ...R USING A SUITABLE MEAN E G TRAN SPALLET IN ACCORDANCE WITH THE REGULATIONS ON THE MANUAL HANDLING OF LOADS IN ORDER TO AVOID ADVERSE CONDITIONS THAT INVOL VE RISK OF ERGONOMIC INJURIES TO THE BACK Th...

Page 25: ...USE OF THE MACHINE placed on a flat floor that ensures stability in relation to the overall dimensions and to the weight IN ORDER TO ENSURE THAT THE AUTHORIZED OPERATORS OCCUPY THE WORK POSITIONS EXPE...

Page 26: ...NE RE ADY FOR THE OPERATION PNEUMATIC CONNECTION THE MANUFACTURER HAS NO LIABILITY FOR INJURY TO PERSONS ANI MALS AND OR THINGS RESULTING FROM FAILURE TO OBSERVE THE WAR NINGS DESCRIBED ABOVE IT IS ST...

Page 27: ...AN ELECTRIC CUTTING DEVICE WITH A SUITABLE DISCONNECTING PROTECTION COORDINATED WITH THE GROUNDING PLANT ANY TYPE OF ELECTRIC MATERIAL USED FOR THE CONNECTION MUST BE SUITA BLE FOR USE MARKED CE IF SU...

Page 28: ...0 5 bar installed in the pneumatic system under the pressure switch It is used to discharge the overpressure of the system if the pressure switch for any problems does not function The intervention of...

Page 29: ...RY 3 40 HS 1 1 2 2 2 3 3 3 4 4 4 4 5 5 5 1 Instruction Manual Clinic 3 40 Compressors IT IS FORBIDDEN TO MANIPULATE EXCLUDE REMOVE AND OR REPLACE ANY SAFETY DEVICE IN THIS MACHINE IT IS ABSOLUTELY FOR...

Page 30: ...ion forbidden and not reasonably foreseeable PROBABILITY Low and accidental STAGE WHERE THERE IS A RISK During use and operation MEASURES TAKEN Safety signs Compulsory use of personal protective equip...

Page 31: ...OD VISIBILITY IT IS COMPULSORY TO REPLACE THE SAFE TY SIGNS DETERIORATED BY REQUESTING THEM TO THE MANUFACTURER AND OR AU THORIZED DEALER RESIDUAL RISK N 2 DANGER OF CRUSHING FOOT DUE TO ACCIDENTAL FA...

Page 32: ...ratures USE AND OPERATION PLACES OCCUPIED BY AUTHORIZED OPERATORS The Authorized operators depending on the type of manual intervention required must take the places shown in Fig 1 THE AUTHORIZED OPER...

Page 33: ...gentilinair com Fig 2 Instruction Manual Clinic 3 40 Compressors PRESSURE SWITCH In Fig 2 is represented the control device to which the authorized operator must refer for the command and control of...

Page 34: ...tart of the machine when the pressure drops to 6 bar and the automatic stop when the pressure reaches 8 bar 4 Interruttore ON I OFF 0 Pos ON I Turns on the machine Pos OFF 0 Turns off the machine 5 Sa...

Page 35: ...MAN0340001 07 2019 35 www gentilinair com CLINIC 3 40 CLINIC DRY 3 40 H CLINIC DRY 3 40 HS 2 2 2 5 5 5 5 3 3 3 4 4 4 1 1 1 Instruction Manual Clinic 3 40 Compressors...

Page 36: ...ne maintenance has been made TURNING ON THE MACHINE THE AUTHORIZED OPERATORS CAN TURN THE MACHINE ON ONLY AFTER THEY HAVE PERFORMED AS A COMPULSORY OPERATION THE CONTROLS DESCRIBED IN The controls tab...

Page 37: ...MAN0340001 07 2019 37 www gentilinair com CLINIC 3 40 CLINIC DRY 3 40 H CLINIC DRY 3 40 HS 18 18 18 12 12 12 Instruction Manual Clinic 3 40 Compressors...

Page 38: ...TO OBTAIN THE INTERVENTION OF MAINTENANCE TECHNICIANS AND OR THE AUTHORIZED DEALER TURNING OFF THE MACHINE At the end of the work cycle perform the turning off of the machine as fol lows 1 Set the swi...

Page 39: ...MAN0340001 07 2019 39 www gentilinair com 15 CLINIC 3 40 CLINIC DRY 3 40 H CLINIC DRY 3 40 HS 18 18 18 12 12 12 15 15 15 7 7 7 Instruction Manual Clinic 3 40 Compressors...

Page 40: ...the switch 18 in pos OFF 0 2 If necessary repeat the operations described in par TURNING ON THE MACHINE INTERVENTION ON THE SAFETY VALVE In the event of a malfunction of the pressure switch and the pr...

Page 41: ...king properly and inserted into the socket Check that any extension cords used and the current outlet are operatio nal Check that the switch of the main po wer supply is functioning and in pos ON I Th...

Page 42: ...ions of use and operationofthemachine throughvarioustypesofintervention adjustments visual inspection recovery levels etc performed by the maintenance tech nician or by the manufacturer s technician a...

Page 43: ...of the tank following the steps below 1 Place under the condensate drain valve a col lection container fit for use 2 Slowly open the valve of condensate drain 12 and wait for a complete exit of compre...

Page 44: ...e the filter 3 and clean it with compressed air REPLACE THE FILTER IN CASE OF EVIDENT WEAR 4 After placing back the filter replace the cover properly 5 Indicate the action on the maintenance card YOU...

Page 45: ...screw the filter D Screw in the new filter Insert the cup in the top cover and turn 45 AF 30 CARTRIDGE REPLACEMENT Press the lever down E turn of 45 the cup F and remove it from the cover G Remove the...

Page 46: ...B Instruction Manual Clinic 3 40 Compressors STERILE FILTER REPLACEMENT Unscrew the cover by hand a Remove the filter B Insert the new filter Screw the cover up until it reaches the mechanic stop Repo...

Page 47: ...ITION Upon demolition of the machine strictly follow the requirements in force in the country of demolition Proceed to the differentiation of the parts that make up the machine accord ing to the diffe...

Page 48: ...te drain no compromise is allowed it is a real float drain that opera tes depending on the rising level of condensate and not of a partial device that can only drain at the end of pumping Core dehydra...

Page 49: ...ce with that governed by the FDA Standard Food and Drug Administration and can be sterilised in autoclave according to conventional criteria DRY TEC DRYING SYSTEM DATA SHEET FLUID Compressed air gener...

Page 50: ...rispettato il corretto processo d inibizione della crescita dei batteri e dei virus fermati dall elemento comunque consigliato sterilizzare il filtro prima di ogni sequenza di lavoro asettico Questi e...

Page 51: ...3 40 Year Details Replacements Hours 1 500 2 Date Signature F2 1000 3 1500 4 Date Signature F2 G1 2000 5 2500 6 Date Signature F2 3000 7 3500 8 Date Signature F2 G1 G2 4000 9 4500 10 Date Signature F...

Page 52: ...etails Replacements Hours 1 500 2 Date Signature F2 F3 1000 3 1500 4 Date Signature F2 F3 G1 2000 5 2500 6 Date Signature F2 F3 3000 7 3500 8 Date Signature F2 F3 G1 G2 4000 9 4500 10 Date Signature F...

Page 53: ......

Page 54: ...Via delle Tezze 20 22 36070 TRISSINO VI Italy phone 39 0445 96 20 00 info gentilinair com www gentilinair com M A N 0 3 4 0 0 0 1 Compressori Clinic 3 40_MAN0340001_07 2019 MAN0340001 07 2019...

Reviews: