• Anti-bacterial filter (option): the output head is already set-up to house an anti-septic
filtering element. It is offered as an optional so as not to increase the purchase cost
of the equipment is not needed, so as to satisfy the more demanding customer without
further transformations. The components of this sterile element are in compliance with
that governed by the FDA Standard (Food and Drug Administration) and can be sterilised
in autoclave according to conventional criteria.
DRY tEC DRYINg SYStEm - Data SHEEt
FLUID:
Compressed air generated bynon-lubricated compressor
Saturated with water vapour
No pulses
SERVICE: ............................................................ continuous, 24/24 hours
INPUT COMPRESSED AIR FLOW RATE: ...................................... 145 sl/min.
OUTPUT COMPRESSED AIR USEFUL FLOW RATE: ........................ 123 sl/min.
COMPRESSED AIR INPUT TEMPERATURE: ...................... from +2°C to +50°C
SUPPLY RATED PRESSURE: .............................................................. 7 bar
DRYER DEW POINT AT ROOM TEMPERATURE OF 25°C: ..................... -19,4°C
PROJECT TEMPERATURE:: ........................................... from +2°C to +50°C
PROJECT PRESSURE: ..............................................................max 10 bar
INSTALLATION: .......................................................................... internal
QUALITY CONSTRUCTION CODE:: ........................................ ISO9001:2008