12
13
EN
EN
FEATURES
C
1
2
5
6
3
4
1) POWER
Power on/off S-Digit multi
+
.
2) ILLUMINATION / SOUND
Press the button short to switch on/off the sound
signal.
Press the button long to switch on/off the illu-
mination
3) HOLD
Press the button short to freeze the current
reading.
4) LASER
Press the button short to switch on/off the laser
beam.
5) CAL
Press the button long to activate the calibration
procedure.
6) MODE
Press the button short to select 5 different display
modes (see chapter „Additional functions“.)
7)
8)
9)
10)
11)
12)
Direction symbol (approching
the 0°/90° position)
Symbol sound on/off
Battery status indicator
Unit symbol N/FT
Unit symbol °/%
Unit symbol mm/m
7
8
9
10
11
12
CHECK AND CALIBRATION
D
The display has an automatic digit inversion for over-
head measurements.
CALIBRATION
Set the instrument down on the same flat and even surface, switch it on and
keep the button
CAL
pressed until the display shows „-0-“. Wait approx. 3 sec.
and press the button
CAL
again until the display shows „-1-“. Turn the unit by
180° and set it down on the same position as before. Wait approx. 3 sec. and
press the button
CAL
again until the display shows „-2-“. Now the calibration is
completed.
CHECK OF CALIBRATION
Set the instrument down on a flat and even surface and switch it on. Wait 3 sec.
and note down the measured value (A). Turn the unit by 180° and set it down
on the same position as before. Wait 3 sec. and note down the measured value
(B). If the deviation between the two values A and B is more than 0,2° the unit
has to be calibrated.
IMPORTANT
Calibration of inclinometer:
- before first use
- before important measurements
- after a hit or drop
- after a large fluctuation in temperature
MEASUREMENT OF INCLINATION
The two arrows in the display indicate in which direction the
instrument has to be moved in order to reach the 0° or 90°
position.
The exact 0° or 90° position is shown by double arrows.
S-Digit multi_DE-EN-FR.indd_Korr..indd 12-13
07.03.2014 10:01:26