36
3.42 “V3” – Display the D Cell Battery Voltage
Displays the three-volt D cell battery pack voltage. Replace the batteries when this voltage is less
than 1.8-volts.
*V3
3V Battery Voltage = 2.93V
*
3.43 “V12” – Display the External Battery Voltage
Displays the external 12-volt battery voltage. Replace or recharge the battery when this voltage
is less than six volts.
*V12
12V Battery Voltage = 12.33V
*
3.44 “WF” - Display Current Wrap Format
Memory “wrapping” means that once the memory has filled, the datalogger will continue taking
readings and overwrite the stored values in a circular fashion (see Section 3.7, “Display Arrays
Forward from User Position”).
When the wrap format is set to 0, logging will stop once the memory becomes full. This is useful
if critical data is stored and it must not be inadvertently overwritten and lost.
When the wrap format is set to 1, logging will continue when the memory becomes full and the
original stored values will be overwritten. With this setting, logging will continue indefinitely
until told to stop with the SP command, the programmed stop time has been reached, or the
battery has fallen to 1.6-volts.
*WF
Logging will not stop when memory is full
*WF0
Logging will stop when memory is full
*WF1
Logging will not stop when memory is full
*