34
SW1 Setting
Channels Accessed
2
3
4
OFF OFF OFF
1-32 (32-channel mode), 1-16 (16-channel mode)
DEFAULT MUX1
OFF OFF ON
33-64 (32-channel mode), 17-32 (16-channel mode)
MUX2
OFF ON OFF
65-96 (32-channel mode), 33-48 (16-channel mode)
MUX3
OFF ON ON
97-128 (32-channel mode), 49-64 (16-channel mode)
MUX4
ON OFF OFF
129-160 (32-channel mode), 65-80 (16-channel mode)
MUX5
ON OFF ON
161-192 (32-channel mode), 81-96 (16-channel mode)
MUX6
ON ON OFF
193-224 (32-channel mode), 97-112 (16-channel mode)
MUX7
ON ON ON
225-256 (32-channel mode), 113-128 (16-channel mode)
MUX8
Table 19 - Daisy-Chain Operation/Channels Accessed
Figure 35 is a schematic representations of the daisy-chain configuration with multiplexers
configured for 32-channels. It shows three multiplexers sharing the same control ports, and a
single cable is used to interconnect them.
Figure 35 - Daisy-Chain Configuration (Does not apply to Model 8600-2)
Reset
Clock
Micro-6000 Datalogger
U5
C4
Multiplexer #1
MUX1
Reset
Clock
Reset
Reset
Clock
Clock
Multiplexer #2
MUX2
Multiplexer #3
MUX3
H
J
H
H
J
J
VW Gages 1-32
VW Gages 65-96
VW Gages 33-64
H
J
H
J
IN
IN
IN
OUT
OUT
SW1 SETTING:
2 3 4
OFF OFF OFF
SW1 SETTING:
2 3 4
OFF OFF ON
SW1 SETTING:
2 3 4
OFF ON OFF