20
Note:
If logging is currently started and the clock is changed, a restart of the scan interval or log
interval table will occur. If the scan interval is fast (e.g., three seconds), logging may need to be
stopped and restarted, depending on whether or not the internal “next time to read” register
becomes unsynchronized with the current time. This is true with both the command line interface
and setting the datalogger clock within LogView.
4.4 “DEFAULT” - Load Factory Default Settings
All stored readings as well as the ID, networking and real-time clock settings are not affected by
this command. User will be asked to verify before executing. Press
Y
to continue, any other key
to abort.
*DEFAULT
This will load the factory default settings!
Are you sure(Y/N)?Y
Restored to factory default settings.
*
LC-2 setup after DEFAULT has been issued:
*S
MS:0 OP:1 UP:1
GT: 1 ZR: 0.00000 GF: 1.00000 GO: 0.00000
Scan interval: 3 second(s).
Logging stopped.
Log intervals disabled.
Monitor mode enabled.
*SS
Signature of RAM1: 58294
Signature of RAM2: 47463
Signature of RAM3: 51545
Signature of RAM4: 15546
Signature of ROM: 4497
Trap count: 0
Network address: 1
Network recognition disabled.
Time format is hhmm.
Date format is Julian.
Standard Temp thermistor selected
*
*WF
Logging will not stop when memory is full
*
*SR
Readings are synchronized to the top of the hour.
*
*RT
8000 readings maximum