35
Line
Description
Manual Sections
1
Status of memory pointers
4.6, 4.26
2
Gauge information
4.8
3
Scan interval setting
4.36
4
Start/Stop status
4.38, 4.39
5
Stop time (optional)
4.38
6
Log interval status
4.13, 4.14
7
Monitor mode status
4.15
Table 6 - S Command Information
4.36 “SCnnnnn” - Set Scan Interval
Enter the scan interval, in seconds. Range of entry is 3 to 86400. Only whole numbers are
accepted. Typing
SC
with no value returns the current setting only
*SC
Scan interval: 60 second(s).
*SC300
Scan interval: 300 second(s).
*
4.37 “SS” - Display System Status
Displays the system status of the datalogger.
*SS
Signature of RAM1: 32819
Signature of RAM2: 15979
Signature of RAM3: 63255
Signature of RAM4: 2197
Signature of ROM: 15283
Trap count: 0
Network address: 1
Network recognition disabled.
Time format is hhmm.
Date format is Julian.
Standard Temp thermistor selected
*