5. MAINTENANCE
Although Model LC-2 Datalogger is designed to operate in field environments, nevertheless
there are some basic maintenance procedures that should be followed to insure maximum
reliability and functionality.
5.1 Keeping the Inside of the Box Dry
The LC-2 datalogger is designed to be splash proof and rain proof but is not designed to be
submersible under water. The LC-2 enclosure lid is sealed by a gasket, which will remain sealed
so long as the lid screws are kept tight. Most important is to make sure that the Hubble connector
around the cable entry is securely tightened so that the internal grommet grips and seals around
the cable. LC-2 models that use a 10-pin connector as a cable entry are equipped with sealing
caps, which must be kept tightened to the connector when the connector is not in use.
Despite all these precautions, the LC-2 may encounter leakage along the cable if the cable is cut,
and/or condensation problems especially in humid environments. In these environments, it is
recommended that the internal desiccant pack be replaced at the necessary intervals to prevent
condensation from corroding or shorting out the internal electronics.
5.2 Cleaning
The outside of the box can be cleaned using a cloth dampened with soap and water. DO NOT
USE ANY TYPES OF SOLVENTS OR SCOURING AGENTS!
The connector sockets can be cleaned using a small stiff brush (small painters brush) dipped in
soap and water. The sockets are water resistant; the internal electronics will not be adversely
affected by them filling with water or other liquids. Be aware however, readings could be
affected by shorting or other effects of an improper connection due to fluids being present in the
connector. Dry connections thoroughly before use.
5.3 Batteries
When the unit is not in use, especially for extended periods of time, the batteries should be
removed to prevent damage due to leakage. The warranty does not cover damage due to battery
leakage. The table below details the approximate operating times for the various types of D cell
batteries that may be used with the Model LC-2.