65
11)
While holding the bulkhead with one wrench, rotate the Swagelok nut with the second
wrench 1-1/4 turns using the marks as a reference point.
Figure 31 - Nut Position After Tightening
J.2 Changing Batteries
It is only necessary to remove the screws holding the lid in place. Do not loosen the Swagelok
fitting.
J.3 Cable Removal/Reinstall
1)
Loosen the Swagelok nut and remove screws holding the lid in place.
2)
Disconnect the leads from the terminal block.
3)
Gently remove cable from Swagelok bulkhead.
4)
Nylon ferrules can be removed from cable by clipping them off. Be careful not to damage the
cable sheath. (Only remove ferrules if not using same sensor, see below.)
5)
If same sensor will be reinstalled, feed the cable back through the Swagelok bulkhead and
attach the bare leads. Finger Tighten and this time only snug the fitting (DO NOT ROTATE
1-1/4 TURNS). The ferrules can be loosened and tightened multiple times on the same cable
if they are not over tightened. Damaged ferrules should be replaced.