40
3.43 V12
Display the external 12V battery voltage. Replace or recharge the battery when this voltage is
less than 6V
*V12
12V Battery Voltage = 12.33V
*
3.44 WF
Display the current Wrap Format. Memory “wrapping” means that once the memory has filled,
the datalogger will continue taking readings and overwrite the stored values in a circular fashion
(see Section 3.7, Dnnnnn).
When the wrap format is set to 0, logging will stop once the memory becomes full. This is useful
if critical data is stored and it must not be inadvertently overwritten and lost.
When the wrap format is set to 1, logging will continue when the memory becomes full and the
original stored values will be overwritten. With this setting, logging will continue indefinitely
until told to stop with the SP command, or the programmed stop time has been reached.
*WF
Logging will not stop when memory is full
*WF0
Logging will stop when memory is full
*WF1
Logging will not stop when memory is full
*
3.45 X
Take and display one reading, but do not store this reading in memory. Useful if interested in
obtaining a reading at the moment, without interrupting or affecting the current logging schedule.
*D
MS:3 OP:4 UP:3
*X
Datalogger#1,2007,11,27,12,32,07,3.10,25.51,9038.122,8055.655,7187.985,9018.434,
---,---,---,---,
685.209,712.577,611.013,683.002,---,---,---,---,15.7,15.7,15.6,15.6,---,---,
---,---,24.8,24.8,24.9,24.8,---,---,---,---
*D
MS:3 OP:4 UP:3
*
Note:
In this example, channels 5,6,7,8 and 13,14,15,16 are disabled.