Operation Messages
ZDL700 |
*
18
9. Operation Messages
E114
Invalid Measurement !
Make another measurement. If further measurement proved to be futile, check staff position and
Inverse Staff setting, check the lighting condition at the staff and stray light, check focusing and
targeting, check if sufficient length of barcode in the field of view.
E115
Temperature Sensor
Error !
Cover the objective telescope with a hand and switch ON the instrument. Hardware communication
failed.
E116
Adjustment Error !
Carry out the adjustment with guided steps, make sure instrument is level and staff is truly vertical in
normal position. Collimation is out of correction range.
E117
BM Change Not Allowed!
Exit to default measurement mode by pressing HEIGHT/DISTANCE key, and change the Benchmark
elevation in the INPUT BM menu mode.
E119
Staff Blocked
Not enough barcode length for measurement.
E120
Image sensor Error!
Contact services.
E121
Adjustment Inverse Staff
Not Allowed!
Check the staff orientation and staff setting.
E123
PtID Change Not Allowed Exit the message by pressing ESC key.
No.
Error message
Counter measure / causes
Operation message
Counter measure / remark
Start Tracking!
Tracking mode starts.
Stop Tracking!
Tracking mode stops.
Tracking Hold!
Press measure key for 3 seconds to restart tracking mode. Tracking will hold after 10 unsuccessful meas-
urements.
Abort Measurement!
Current measuring process is terminated.