10.2
Adjusting the Self-Levelling Accuracy
011294_001
X
In Adjustment Mode the X-axis LED indicates changes
to the X-axis.
011295_001
Y
The Y-axis LED indicates changes to the Y-axis.
1.
Turn off the power.
2.
Put the Zone60 DG in an upright position.
3.
Press and hold both the Up and Down arrow buttons.
4.
Press the Power button. The X-axis calibration screen appears. The Zone60 DG is
now in calibration mode.
☞
In calibration mode, the LED does not blink and the laser head continues to rotate.
An hour-glass indicates that the Zone60 DG is levelling.
When entering Calibration mode, the X-axis calibration screen appears:
007733_001
1.
When the hour glass has disappeared, indicating that the Zone60 DG has levelled,
check both sides of the X-axis.
2.
Press the Up and Down arrow buttons to bring the plane of laser light to the spe-
cified level position.
☞
Each step represents approximately 2 arc seconds of change. Therefore, 5 steps
equal approximately 1.5 mm at 30 m (1/16" at 100’).
3.
Press the Grade button to accept the adjusted position and to switch to the Y-axis
calibration screen.
Description
Entering calibration mode
step-by-step
Calibrating the X-axis
step-by-step
Accuracy Adjustment
55