6
2. EM38-MK2 OPERATING INSTRUCTIONS - Expanded Method
The following procedures are used to prepare the EM38-MK2 for survey operation. The
suggested time interval between the various procedures are typical values; under adverse
conditions they may have to be carried out more often and under good conditions less often, as
dictated by survey experience. Since the procedure is similar for both coil separations, only the
procedure for the 1 m coil separation will be described
2.1
Battery Test
The battery test should be carried out at the beginning of each day or when the battery voltage is
suspected of being low. Check the battery voltage as follow: set the MODE switch to BAT.
In the case of a good battery meters will read over 720 units. If meters read below 720 units
replace the battery. To remove the battery, undo the two thumb screws holding the cover of the
battery compartment. A Mallory MN1604 9-volt Alkaline battery or equivalent gives about 5
hours of continuous operation. For longer service (about 12 hours) we recommend a
ENERGIZER L522 9-VOLT Lithium battery.
Geonics also provides as an option, an external battery pack that give about 25 hours of
continuous service. For more details see section 10.
2.2
Initial Inphase Nulling
The initial inphase nulling should be carried out at the beginning of the day at the first survey
station. As described in Technical Note TN-6 the EM38 measures ground conductivity by
inducing very small electrical "eddy" currents in the ground and measuring the magnetic field
that these currents generate. A small transmitter coil located at the front of the EM38-MK2 is
used to generate the time-varying primary magnetic field which induces the eddy currents in the
ground, and a small receiver coil located at the end (middle for the 0.5 m separation) measures
both this strong primary magnetic field and the much smaller secondary magnetic field arising
from the eddy currents in the ground.
It is the task of the receiver electronics to measure the very small signal from the eddy currents in
the presence of the much larger signal arising from the primary magnetic field. To facilitate this
measurement an internally generated signal is used to cancel or "null" the large primary signal so
that it does not overload the electronic circuitry. Because of the high sensitivity of the EM38-
MK2 it is advisable to remove all metal objects from wrists, fingers, neck and pockets when
making this adjustment (sensitivity to metal objects, which is discussed further in Section 2. is
greatest near the coils at either end of the instrument).
To null the EM38-MK2, lift the instrument to a height of about 1.5 meters (5 feet) above the
ground and place in the horizontal dipole mode of operation (Fig. 1). Set the Mode switch to the
1 m position and null the I/P meter to indicate zero by 1 m control. The EM38-MK2 is
satisfactorily nulled when the instrument at a height of 1.5 meters the I/P meter reads
approximately zero (± 10 mS/m).
Summary of Contents for EM38-MK2
Page 7: ...5 Fig A...
Page 23: ......
Page 24: ......
Page 25: ......
Page 26: ......
Page 27: ......
Page 28: ......
Page 29: ......
Page 30: ......
Page 31: ......
Page 32: ......
Page 33: ......
Page 34: ......
Page 35: ......
Page 36: ...34...
Page 37: ...35...
Page 38: ...36...
Page 39: ...37...
Page 44: ...42...