Instructions for Use
Ambistik PFT-CPET
Serial no:
2024xxxxx / 2025xxxxx
Page 56
Version: 1 | Release date: 31 January 2022
12.2
Installation and Operating Conditions
The following conditions supplement chap. 2.2 "Intended Use"
and must be observed to maintain the properties of the Ambistik
PFT / Ambistik CPET guaranteed by the manufacturer and for
the safety of the patient and operator.
Storage / Transport:
min.
max.
Temperature:
-10 °C
+ 50 °C
Relative air humidity:
10 %
95 %
non-condensing
Operation:
min.
max.
Environmental temperature:
+10 °C
+35 °C
Avoid extreme
fluctuations in
temperature!
Relative air humidity:
20%
90 %
non-condensing
(min. 30 % for synthetic flooring)
Atmospheric pressure:
700 hPa
1100 hPa
Setup:
mobile
Environment / room:
not an explosive or flammable environment!
not near an MRI!
Operating Mode:
Continuous operation