Instructions for Use
CPET
Version: 4 | Release date: 01 April 2022
page 31
3.1.3.3.3
Start of the Recovery Phase
1.
Start recovery.
For this:
Press the
"Recovery" [1]
button.
In the recovery phase, the physical exercise for the patient is
reduced to a low level. During the recovery phase, the measured
values continue to be recorded continuously. This serves, among
other things, to detect stress-induced side effects.
With the start of the recovery phase, the face mask can be
removed together with the sensor. During this phase, which lasts
about 5 to 7 minutes, the monitoring of blood pressure and
oxygen saturation continues.
During the recovery phase, the events can be used to record
additional data via the toolbar (see chap. 3.1.3.5).
1