Instructions for Use
CPET
Version: 4 | Release date: 01 April 2022
page 71
5
Settings
5.1
Measurement CPET
Setting
Function
MVV from FEV1 Adjust how the MVV should be calculated. In this
example 35 x FEV1 (measured)
Mask
deadspace (ml)
Configure the dead space value of the mask or
mouthpiece being used for testing
Tube length
Adjust the length of the tube
Blood pressure
confir..
If activated the result of automatic BP measurement
need to be confirmed
Insp.
Deadspace
Setting of inspiratory dead space. Need to be adjusted if
apparatus in front of flowsensor will be added to manage
different source of inspired gas concentration.
FIO
2
If activated FiO
2
will be set to 20.94
FiCO
2
If activated, FiCO
2
is measured during inhalation.
Enable logging
If activated, the raw data of the CPET measurement will
be saved additionally.