Troubleshooting

What 
happened 

Causation Correction

The indicator 
turns red.  

The batteries have 
run out of power.

Replace the bat-
teries.

The mesh cap is 
dirty.

Clean the mesh 
cap. If the 
problem persists, 
please replace 
with a new mesh 
cap.

The device 
does not work 
when it is 
turned on 

No power.

Check the batter-
ies and replace 
them if neces-
sary. Pay atten-
tion to polarity.

The indicator 
is green but 
no aerosolis 
produced or 
the delivery is 
slower than 
usual.

The medication cup 
is empty.

Fill medication 
cup and turn on 
the nebulizer.

The mesh cap is 
broken.

Replace with a 
new mesh cap.

The medication cup 
or the mesh cap 
is dirty.

Clean those parts 
or replace mesh 
cap.

There is 
medication in 
the medication 
cup and the 
ON/OFF but-
ton has been 
pressed, but 
no aerosols 
are being 
produced.

No power supply.

Check the batter-
ies and replace 
them if neces-
sary. Pay atten-
tion to polarity.

The mesh is 
covered with liquid.

Check if the 
mesh cap is cov-
ered with liquid. If 
yes, shake off the 
liquid and then 
try again.

Summary of Contents for Micromesh GT-7002

Page 1: ...GB Nebulizer with Micromesh Technology Instruction manual GT 7002...

Page 2: ...on your purchase By purchasing the Geratherm micromesh nebulizer you have acquired a quality product by Geratherm Please read these instructions for use carefully to use the device without problems a...

Page 3: ...se Operation 30 Battery Replacement 30 Operating Instructions 30 Important Information before Use 30 Unit Assembly and Usage 31 Inhalation with Mouth Piece and Masks 33 Cleaning and Disinfection 34 Be...

Page 4: ...the section entitled Cleaning and Disinfection of Nebuliser Parts Accessories Only silicon masks which are available from specialist suppliers can be disinfected by boiling or by means of vaporisation...

Page 5: ...adult children 2 AA Alkaline batteries Mouth piece Mesh cap adapter for masks and mouth piece Medication cup Indicator green usage red battery needs to be replaced green and red reset the vibration r...

Page 6: ...detached from the device may result children choking from inhaling or swallowing The accessible materials used in the device will not cause the potential allergic reactions to skin Do not try to modif...

Page 7: ...ensure efficient nebuliza tion The device is ideally suitable for inhalation at home or when travelling It guarantees highly effective fast acting inhala tion treatment for patients of all ages Carry...

Page 8: ...y until a click is heard Note Remove the batteries if not using the device for three months or longer Discard the used batteries according to your local regula tions Important Information Before Use E...

Page 9: ...The maximum capacity of the medication cup is 6 ml Please note that the medi cation must not exceed this maximum level marking 6 ml 3 Slide the mesh cap into the unit align the 2 knobs with the cutou...

Page 10: ...elivery mode The nebulizer provides you with two delivery modes accord ing to your needs automatic or manual Automatic delivery of aerosol 1 Press and release the ON OFF button The unit turns on and s...

Page 11: ...for two to three sec onds before breathing out This allows the medication to settle into the respiratory system 4 Continue the treatment until the medication has been used up about 10 to 15 minutes 5...

Page 12: ...ise direction 2 Remove the mesh cap by gently pressing the flap and lifting up in the direction indicated the arrows 3 Discard any remaining medications pay attention to statutory provisions Cleaning...

Page 13: ...ng with a cloth moistened with medical alcohol Ethanol 75 dilution with water 3 Let them air dry on a clean paper towel Note Only silicon mask can be disinfected by boiling PVC mask just can be disinf...

Page 14: ...e first time you need to reset the vibration rate 1 Pour water into the medicine cup pay attention to maxi mum marking 2 Attach the mesh cap to the nebulizer unit 3 Take the batteries out 4 While hold...

Page 15: ...replaced after continuous use Accessories of the nebulizer micromesh involve mesh cap mesh adapter for masks and mouth piece PVC masks for adult and children silicon masks for adult and chil dren and...

Page 16: ...o aerosolis produced or the delivery is slower than usual The medication cup is empty Fill medication cup and turn on the nebulizer The mesh cap is broken Replace with a new mesh cap The medication cu...

Page 17: ...protected gainst dripping water Automatic function Automatic power shut off if system idles for 15 minutes Power consumption Normal mode less than 1 5 W Standby mode less than 0 3 mW Quality standard...

Page 18: ...onsult instructions for use Use by expiry date for masks Batch code for nebulizer Manufacturing date of masks Manufacturer Serial number Caution consult acompanying docu ments Do not dispose of as hou...

Page 19: ...able manner Please remove the batteries before disposing of the de vice unit Do not dispose of old batteries with your house hold waste but at a battery collection station at a recycling site or in a...

Page 20: ...hat is caused through improper handling through use otherwise than in accordance with the instructions in this manual or through unauthorised maintenance attempts Your local dealer cannot declare this...

Reviews: