12.
Spray a clean wipe with disinfectant cleaner and wipe the front air return grill.
13.
Close the front window. Clean the glove sleeves and gloves and replace on the front
window.
14.
Open the front window and repeat steps 1-13, without removing the sleeves and
gloves, with sterile water to rinse away any disinfectant.
15.
Turn on the LFGI and close the front access window.
16.
Spray 70% IPA directly on all surfaces of the rest of the work area, under sliding
work trays, the inside of the front access window, gloves and sleeves.
17.
Allow the LFGI to run for 10 minutes before use.
If the LFGI is under negative pressure all cleaning steps must be performed as
described above with the viewing panel closed and through attached sleeves/gloves.
GERMFREE can provide an isolator cleaning tool to assist you in this function,
contact us at 800-888-5357 for more information.
Cleaning the LFGI between preparations- Light
Sanitizing the LFGI work area between preparations is recommended to reduce the chances
of cross contamination. 70% Isopropyl Alcohol is the most recommended agent for this
task.
1. Open only the outer door of the airlock. Spray a clean wipe with alcohol and wipe the top and
back HEPA filter covers.
2. Spray the remaining surfaces, including the tray with alcohol and wipe with overlapping
strokes.
3. Leave the spray bottle in the airlock and close the outer airlock door. Run a purge cycle.
4. Open the inner airlock door and remove the spray bottle.
5. Spray all vertical surfaces in the work area with alcohol and wipe from top to bottom with
clean wipes, overlapping strokes moving left to right.
6. Spray the air grills and work surface with alcohol and wipe from left to right with overlapping
strokes moving back to front.
7. Spray the sleeves and gloves and rub together, then wipe off.
8. Spray the same surfaces lightly and allow to air dry.
43
Summary of Contents for LFGI-3USP
Page 1: ...Laminar Flow LFGI CAI or CACI User Manual ...
Page 33: ...33 ...
Page 51: ...SECTION IV UNIT DIAGRAM Made in the USA and available for purchase on the GSA 51 ...
Page 55: ...55 ...
Page 59: ...59 ...
Page 63: ...Fig 2 Air recycle plenum connection tip removed 63 ...
Page 85: ... 863 248 2800 www sensocon com 20 Installation and Operation Manual Series A3 A4 ...