12
2
Before lifting the patient, make sure the spade is securely
supported and locked by the quick coupling.
(See Fig. 6)
3
Loosen the spade seat by pressing the ejector and lift the
quick coupling upwards, holding the spade seat with the
other hand.
(See Fig. 7)
Emergency Stop
If, for any reason, the
Walker
aid does not respond to the control
buttons, you can quickly stop lift movement by pressing the
button on the battery and disconnect it.
(See Fig. 1)
Emergency High/Low
(4 steps)
If, for any reason, the
Walker
aid does not respond to the control
buttons, raise or lower the
Walker
aid by using the Emergency
High/Low.
1
Unplug the hand control.
(See Fig. 2)
2
Control the
Walker
aid by pressing a blunt, thin object into
the holes at the panel. (i.e. a pen).
(See Fig. 3)
3
Press it into the hole next to the “up arrow” in order to raise
the
Walker
aid.
4
Press it into the hole next to the “down arrow” in order to
lower the
Walker
aid.
Contact your local ArjoHuntleigh representative before re-using
the
Walker
aid.
CAUTION
Do not use an object with sharp tip when
pressing in the holes.
NOTE
A continuous beeping sound is heard during action
“up” and “down”.
Fig. 6
Fig. 7
Fig. 1
Fig. 2
Fig. 3