|
13
|
IAB Insertion / CARDIOSAVE Operation
|
Cardiac Assist
|
Initial Set-up using a Conventional IAB (continued)
Pressing the START key
•
Automatically purges and fills IAB
•
Automatically selects most appropriate
lead and trigger
•
Automatically sets inflation and deflation timing
Verify Aug. Alarm is set 10mmHg below
patient’s augmented diastolic pressure
(approx. 3 minutes after initiation of assist).
1
2
2
1
2
1
1
2
1
2
He
Inflated
Deflated
Balloon Pressure
116
Aug Alarm
Augmentation
IAB Frequency
1:1
106
mmHg
mmHg
Pressure
130
80
30
Arterial
Source: Fiber-Optic
102
52
(84)
MEAN
Inflation
Deflation
Auto
0.0
Trigger: ECG
bpm
Rate
80
Lead: Auto - II
ECG
0.5
mV
1
2
2
1
2
1
1
2
1
2
He
Inflated
Deflated
Balloon Pressure
116
Aug Alarm
Augmentation
IAB Frequency
1:1
106
mmHg
mmHg
Pressure
130
80
30
Arterial
Source: Fiber-Optic
102
52
(84)
MEAN
Inflation
Deflation
Auto
0.0
Trigger: ECG
bpm
Rate
80
Lead: Auto - II
ECG
0.5
mV
6
Verify
AUTO
Press
START
7
CARDIOSAVE SET-UP