22
|
IAB Insertion / CARDIOSAVE Operation
|
Cardiac Assist
|
•
Minimize length of pressure tubing
•
Use
only
low compliance pressure tubing
•
Elevate flush bag at least 3
’
(91.44cm)
above transducer
•
A 3cc/hour continuous flow through
inner lumen is recommended
•
If inner lumen becomes damped
-
Aspirate and discard 3cc of blood
-
If
unable
to aspirate blood, consider
inner lumen clotted, cap lumen, provide
alternate pressure source
-
If
able
to aspirate blood, fast flush to
clear pressure tubing
for
at least
15 seconds
(with IABP on Standby)
•
Do not
sample blood from inner lumen
Proper Care of Inner Lumen
≤ 5' Preferred
8' Maximum
3' Minimum
CARDIOSAVE
OPERATION