23
|
IAB Insertion / CARDIOSAVE Operation
|
Cardiac Assist
|
Changing the Helium tank
Note: Once the helium alarm sounds, there are
24 Autofills remaining in tank.
Install fresh helium tank.
1a
Grasp both sides of
helium tank panel and
1b
Pull out to open.
Grasp helium tank and
slowly slide drawer out.
3a
Remove helium tank
knob from Holder and
3b
Attach to left end
of helium tank, then
fully turn clockwise
to close.
Slowly loosen yoke
T-handle counter-
clockwise (some helium
may escape).
Remove helium tank.
Slowly open helium
tank knob counter-
clockwise (listen for
any escaping helium).
Slide helium tank drawer
in and replace helium
tank panel.
Fully tighten yoke T-handle
clockwise.
Replace washer,
if available.
Verify full helium level
via icon on Monitor Display.
9
10
8
7
11
He
4
5
2
1a
1b
3b
3a
6
CARDIOSAVE
OPERATION