27
|
IAB Insertion / CARDIOSAVE Operation
|
Cardiac Assist
|
Alarms
Probable Cause
Corrective Action
IAB catheter or extender
tubing is disconnected.
Reattach IAB, press START.
IAB Catheter Restriction
IAB Disconnected
Probable Cause
Corrective Action
Restriction in IAB
catheter or tubing.
Relieve restriction
if possible, press START.
Membrane has not
completely unfolded.
Manually inflate and deflate
IAB.
IAB remains in sheath.
Check markings on IAB and
if IAB has not exited sheath,
refer to IFU to reposition
sheath relative to IAB catheter.
TROUBLESHOOTING