31
|
IAB Insertion / CARDIOSAVE Operation
|
Cardiac Assist
|
Alarms
Fiber-Optic Sensor Module Failure
Probable Cause
Corrective Action
There has been a failure
of the internal Fiber-Optic
Sensor Module in the IABP.
If a fiber-optic IAB is NOT in
use, continue normal IABP
use.
If a Maquet/Datascope
fiber-optic IAB is in use,
replace IABP with another
Maquet/Datascope IABP that
supports the fiber-optic IAB.
If replacement IABP is not
available, provide alternate
A.P. source (i.e.: radial).
Contact Maquet Service for
Fiber-Optic Sensor Module
repair.
Probable Cause
Corrective Action
There is a failure in
communication of the
fiber-optic sensor signal
with the IABP.
Unplug Fiber-Optic Sensor
Connector and reconnect.
Relieve any visible kinks in
orange Fiber-Optic Cable.
If problem persists, discon-
nect Fiber-Optic Sensor
Connector and provide
alternate A.P. source
(i.e.: radial).
Fiber-Optic Sensor Failure
TROUBLESHOOTING