35
|
IAB Insertion / CARDIOSAVE Operation
|
Cardiac Assist
|
TRANSPORT
CONSIDERATIONS
Inserting the Pump Console into the Hospital Cart
Grab handles located on
top and front of console,
then lift into Hospital
Cart.
Squeeze latches and attach
to display mount, then
release the latches. Ensure
monitor is securely attached.
Squeeze latches located
below Monitor and lift
to remove from Pump
Console.
Squeeze latch below
handle. Push straight
down until wheels
retract and handle is
fully collapsed.
4a
Push and hold button on console
to unlock monitor mount.
4b
Push down to lock into place,
then release button.
Grab handle and slowly slide
console into Hospital Cart
until it locks into place.
Note: An audible click will
be heard when console is
locking into cart.
To ensure console was successfully installed into cart, plug
Hospital Cart power cord into a compatible grounded AC
receptacle. Confirm AC operation by presence of AC Plug Icon
in the Battery Icon Display Area. If AC Plug Icon is not present:
- Release latch on the Hospital Cart located below Pump Console
- Grab the handle and slowly slide Console out approximately
one quarter of the way
- Repeat steps 6-7 to ensure Pump Console was successfully
installed into Hospital Cart
5
Unable to Update Timing
Check IAB Catheter
Check IAB Catheter
!
!!
!!!
Poor Signal Quality
System Failure
No Trigger
Poor Signal Quality
!!!
!!
Hold Power Button for 2 Seconds to Shutdown
Poor Signal Quality
!!
No Trigger
!!!
Check IAB Catheter
!!!
1
2
1
2
min.
< 5
1
2
1
2
1
2
1
22
1
He
INFLATED
DEFLATED
Balloon Pressure
Time in Standby: 22 min
112
Aug Alarm
Augmentation
Off
IAB Frequency:
1:1
Off
mmHg
mmHg
Pressure
130
80
30
Arterial
- Not Zeroed
Source: Transducer
96
76
(84)
MEAN
Inflation
Deflation
4.0
R-Wave
Auto
- LL Faulted
Lead: Auto - II
ECG
0.5
mV
125
Trigger: Pacer V
bpm
Rate
7
6
3
2
1
4b
4a