37
|
IAB Insertion / CARDIOSAVE Operation
|
Cardiac Assist
|
Connecting an Arrow IAB/IABP
to a Maquet/Datascope IABP
Transferring Facility
•
This patient will have an Arrow IAB connected to an Arrow IABP
•
Before leaving facility, locate IAB catheter extender tubing supplied
in Arrow IAB box, which connects an Arrow IAB to a Maquet/
Datascope IABP
•
Take this IAB catheter extender tubing on transport with patient,
for use when arriving at receiving facility
Receiving Facility
•
When arriving at receiving facility, remove current IAB catheter
extender tubing that connects an Arrow IAB to an Arrow IABP
•
Connect appropriate end of IAB catheter extender tubing (that
was brought from transferring facility) to Arrow IAB, then connect
male luer fitting of IAB catheter extender tubing to back of
Maquet/Datascope IABP
•
Set-up Maquet/Datascope IABP per abbreviated instructions
on page 11 of this Quick Reference Guide
Connecting a Maquet/Datascope
IAB/IABP to an Arrow IABP
Transferring Facility
•
This patient will have a Maquet/Datascope IAB connected to a Maquet/
Datascope IABP
•
Before leaving facility, locate Arrow Pump Adapter (APA) that connects
a Maquet/Datascope IAB to an Arrow IABP (may be supplied in Maquet/
Datascope IAB box or separately)
•
Take the APA on transport with patient, for use when arriving at receiving
facility
Receiving Facility
•
When arriving at receiving facility, place Maquet/Datascope IABP on
Standby and disconnect IAB catheter extender tubing from back of IABP
•
Connect Arrow Pump Adapter (APA) to male luer fitting of Maquet/
Datascope IAB catheter extender tubing and connect to Arrow IABP
•
Adjust volume setting on Arrow IABP, according to Operating
Instructions, to match IAB catheter volume
CONNECTING TO
OTHER IABP SYSTEMS