38
|
IAB Insertion / CARDIOSAVE Operation
|
Cardiac Assist
|
Pneumatic Module Leak test
Press and hold
A.P. OUTPUT VENT
button.
Press and release
POWER button
to turn IABP ON.
This test measures pneumatic leak rate(s) of the Pneumatic Module
and is recommended to be performed before or after each IABP use.
WARNING:
Pneumatic Module
Leak Test MUST NOT be performed
with the pump connected to a patient’s IAB.
•
Press and release POWER button while continuously holding
A.P. OUTPUT VENT button until SPECIAL ACTIVATION MENU
appears on Touchscreen.
•
Press PNEUMATIC MODULE LEAK TEST key, then press
START PNEUMATIC MODULE LEAK TEST key.
•
Using a non-locking luer cap, tightly plug Pneumatic Module outlet
when message PLEASE PLUG IAB PORT appears on Touchscreen
in INSTRUCTIONS field.
•
STATUS field will display pneumatic tests that are currently
being executed. When tests are complete (in approx. 3 minutes),
the message PIM LEAK TESTS COMPLETE will be displayed.
•
If system passes test, the message PASS in green will be displayed
in RESULTS field. Remove non-locking luer cap and press and hold
POWER button for 2 seconds to exit SPECIAL ACTIVATION
MENU. If IABP therapy is being started, then proceed with set-up
of pump.
•
If system fails test, the message FAIL in red will be displayed
in RESULTS field. Check to ensure that non-locking luer cap is tight,
then repeat leak test. If test fails again, contact Maquet Service.
PNEUMATIC MODULE
LEAK TEST