1.2.2 POWER SUPPLY
WARNINGS!
The power cord must be connected only
to a properly grounded AC electrical
outlet to avoid the risk of electrical shock.
If only built-in batteries are mounted
inside the patient unit, then a high priority
alarm is activated if one or more of these
batteries are missing.
CAUTIONS:
Do NOT use antistatic or electrically
conductive tubing with this system.
Avoid contact with external electrical
connector pins.
1.2.3 FIRE HAZARD
WARNINGS!
Keep the system and its gas hoses clear
of all ignition sources.
Do not use the system with worn or
frayed hoses or hoses that have been
contaminated by combustible materials
such as grease or oil.
Oxygen-enriched gas is extremely
flammable: if you detect a burning odor,
disconnect the oxygen supply, mains
power and remove the batteries.
Make sure that both the mains power
outlet and the power supply connector
are accessible.
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
11
|
Introduction
| 1 |
Summary of Contents for Maquet SERVO-s VENTILATOR SYSTEM V7.0
Page 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Page 2: ......
Page 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Page 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Page 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Page 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Page 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Page 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Page 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Page 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Page 183: ...x ...
Page 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...