6.12 BACKUP VENTILATION
Pressure support/
CPAP, NIV PS
Apnea
Pressure control
Backup ventilation is available in all support
modes.
For apnea alarm setting range, see section
The minimum backup pressure level is
5 cmH
2
O.
See the tabulated breathing parameter settings
6.12.1 FUNCTIONAL DESCRIPTION
Backup ventilation switches Pressure Support
and CPAP to Pressure Control.
1
2
3
The following parameters are set:
1.
PC above PEEP (cmH
2
O) in backup
ventilation, Pressure Support
2.
Resp.Rate (b/min) in backup ventilation.
3.
I:E / Ti (s) in backup ventilation (depending
on configuration).
NO PATIENT EFFORT
In case of apnea, the ventilator will switch to
backup ventilation according to the backup
settings at the end of the set apnea time.
The Apnea time is set in the Alarm Profile
window.
Backup is indicated in the active Ventilation
mode touchpad and the alarm No patient effort
is displayed on the screen.
If the patient triggers a breath, the ventilator
automatically switches back to the supported
mode.
110
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
| 6 |
Ventilation, modes and functions
|
Summary of Contents for Maquet SERVO-s VENTILATOR SYSTEM V7.0
Page 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Page 2: ......
Page 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Page 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Page 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Page 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Page 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Page 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Page 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Page 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Page 183: ...x ...
Page 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...