EXPIRATORY HOLD
SVS-5032_XX
Expiratory and inspiratory valves are closed
after the expiration phase is completed, for as
long as the fixed key is depressed, up to a
maximum of 30 seconds. Expiratory hold
provides an exact measurement of the end
expiratory pause pressure. It can be used to
determine total PEEP and, together with
inspiratory hold, static compliance. The
dynamic pressure is shown on the PEEP
numerical value.
INSPIRATORY HOLD
SVS-5031_XX
Inspiratory hold is activated by manually
pressing the fixed key. The maximum time is
30 seconds. The inspiratory and expiratory
valves close after inspiration. This function can
provide an exact measurement of the end
inspiratory lung pressure. It can be used
during x-ray or to determine Plateau pressure,
or, together with expiratory hold, calculate
static compliance.
Read about alarm settings on page 126.
See also Set Ventilation Mode on page 52.
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
117
|
Ventilation, modes and functions
| 6 |
Summary of Contents for Maquet SERVO-s VENTILATOR SYSTEM V7.0
Page 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Page 2: ......
Page 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Page 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Page 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Page 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Page 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Page 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Page 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Page 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Page 183: ...x ...
Page 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...