1.2.4 GASES
CAUTION: The system is not intended to
be used with any anesthetic agent.
Important:
Supplied gases shall meet the
requirements for medical grade gases
according to applicable standards.
Maximum levels:
Air
H
2
O < 7 g/m
3
Oil < 0.5 mg/m
3
Chlorine: must not be detectable
1
Oxygen
H
2
O < 20 mg/m
3
Oil < 0.3 mg/m
3
Note: For devices with serial numbers
below 25000 that are updated to version
7.0, the maximum inlet gas pressure will be
reduced (see the Technical data chapter on
1.2.5 AUXILIARY EQUIPMENT
CAUTIONS:
Accessories, supplies, and auxiliary
equipment used with the ventilator
should:
- be recommended by MAQUET
- meet EN/IEC 60601-1standards
- meet IEC standards as a whole system
If a scavenging system (i.e. gas
evacuation) is connected to the ventilator,
it must conform to ISO8835-3 guidelines
for subatmospheric pressure and
induced flow.
Measurements of parameter values that
have been processed by auxiliary
equipment:
- may be inaccurate if equipment not
authorized by MAQUET is used
- should be discounted if they conflict
with information on the ventilator
screen
- must not substitute for therapeutic or
diagnostic decisions
Note: Applied parts, i.e. equipment making
physical contact with the patient, comprise
nebulizer patient unit and cable and the
Ventilator Breathing System described in
System Flow Chart, Ventilation, Patient
Connection, part no. 66 92 522.
1. If the compressed air is generated by a liquid ring compressor there is a potential risk of chlorine in the supplied
air.
12
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
| 1 |
Introduction
|
Summary of Contents for Maquet SERVO-s VENTILATOR SYSTEM V7.0
Page 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Page 2: ......
Page 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Page 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Page 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Page 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Page 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Page 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Page 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Page 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Page 183: ...x ...
Page 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...