7.1.1 ALARM OUTPUT CONNECTION
OPTION
If your system is equipped with the alarm
output connection option, high and medium
priority alarms can be transferred to an
external signal system. The alarm output signal
is active as long as the audio alarm is active
on the ventilator.
WARNING! Never leave the patient
unattended; the external alarm is designed
to alert those already in attendance.
CAUTION: The alarm output is a
nonguaranteed alarm according to IEC
60601-1-8 and it is recommended that
users establish a Pre-use check routine for
this application.
7.1.2 VISUAL ALARM DISPLAY
SVS-6002_XX
When an alarm is activated, the following
information is supplied on the screen.
1.
A text message explaining the cause of
the alarm flashes in the alarm message
area. The alarm with highest priority is
displayed first.
2.
The corresponding measured value or set
value box flashes and an arrow points at
the exceeded limit.
Note: Two bells in the alarm message area
indicate that more than one alarm is
activated.
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
121
|
Alarms
| 7 |
Summary of Contents for Maquet SERVO-s VENTILATOR SYSTEM V7.0
Page 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Page 2: ......
Page 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Page 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Page 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Page 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Page 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Page 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Page 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Page 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Page 183: ...x ...
Page 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...