7.2.2 RESPONDING TO ALARMS
To respond to a High or Medium priority alarm:
1.
If desired, press the Audio Pause fixed
key for less than two seconds to silence
the alarm for two minutes.
2.
Take action to resolve the alarm condition.
3.
Press the Audio Pause key to reset the
latched high priority alarm and clear the
message from the screen.
The Audio Pause key is identified by the
symbol.
To respond to a Low priority alarm:
1.
If desired, press the Audio Pause key for
less than two seconds to reset the alarm
even if the alarm condition remains.
2.
Take action to resolve the alarm condition.
The alarm is automatically reset once the
alarm condition ceases.
Note:
The following Medium Priority alarms
display an Audio off? message when
activated:
Air Supply Pressure: Low
O
2
Supply Pressure: Low
Battery Operation
For these alarms, you can silence the audio
signal even if the alarm condition is not
resolved. However, the system will
eventually reactivate the alarm.
Using the Audio Pause Key
SVX-5098_EN
Pressing the Audio Pause fixed key for less
than two seconds has the following results:
Active alarms are silenced for two minutes.
A crossed bell symbol along with the time
remaining in the silent period are displayed
in the message area.
Each press of the Audio Pause key restarts
the two minute silent period from when the
user last pressed the key.
Latched alarms are reset if the alarm
condition has ceased.
Note: The No battery capacity alarm and
technical alarms cannot be silenced.
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
123
|
Alarms
| 7 |
Summary of Contents for Maquet SERVO-s VENTILATOR SYSTEM V7.0
Page 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Page 2: ......
Page 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Page 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Page 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Page 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Page 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Page 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Page 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Page 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Page 183: ...x ...
Page 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...