7.2.3 PRE-SILENCING ALARMS
To silence most alarms (active and inactive)
for two minutes, press and hold the Audio
Pause fixed key for more than two seconds.
This action has the following results:
1.
All alarms, active and inactive, are
silenced for two minutes from the time the
key was pressed.
2.
A double crossed bell symbol along with
the time remaining in the silent period are
displayed in the message area.
3.
Pressing the Audio Pause key again for
less than two seconds will now restart the
silent period for two additional minutes
from when the user last pressed the key.
4.
Latched alarms are reset if the alarm
condition has ceased.
SVX-5098_EN
When the alarms are pre-silenced, pressing
and holding the Audio Pause key again for
more than 2 seconds will reactivate all audible
alarms.
124
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
| 7 |
Alarms
|
Summary of Contents for Maquet SERVO-s VENTILATOR SYSTEM V7.0
Page 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Page 2: ......
Page 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Page 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Page 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Page 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Page 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Page 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Page 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Page 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Page 183: ...x ...
Page 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...