9.3 MEDIUM PRIORITY ALARMS
Remedies
Possible causes
Alarm Message
Check the gas supply lines. Perform a
Pre-use check.
Contact a service technician.
Air supply pressure above 6.0 kPa x
100 (87 psi)
Air supply pressure at gas inlet is too
high.
Air supply pressure:
High
Check and connect gas supply lines.
Perform a Pre-use check.
Air supply pressure below
2.0 kPa x 100 (29 psi)
Air supply pressure at gas inlet is too low.
Gas supply line disconnected.
Note:
This alarm can be permanently
silenced (Audio off) when activated.
Air supply pressure:
Low
Contact a service technician.
Technical problems (hardware or
software) with the external alarm
function.
Alarm output
connection error
Check the connection to AC power.
AC power interrupted.
Battery operation
Check the alarm limits.
The persistent memory has corrupt
contents.
Check alarm limits
Check default alarm limits.
Contact a service technician.
Problems in internal memory for default
alarm limits.
Check default alarm
limits
Perform a Pre-use check.
Expiratory cassette has been exchanged
during operation. Pre-use check not
performed after exchange.
Exp. cassette
exchanged
Change ventilator settings.
Combination of settings exceeds the
allowable inspiration flow range.
Inspiratory flow
overrange
Check fan operation.
Check the operating temperature.
Clean the fan filter in the patient unit.
Temperature inside the ventilator is too
high.
Internal
temperature: High
Check patient and breathing system.
Check mask/ prongs size and patient fit.
Check the helmet leakage.
Leakage too high. The mask / prongs /
helmet may not be adjusted properly for
the patient or may be the wrong size.
Leakage out of
range (A medium
priority alarm when
NIV disconnect flow
is disabled). See
section High priority
alarms.
Check the gas supply lines.
Perform a Pre-use check.
Contact a service technician.
O
2
supply pressure above 6.0 kPa x 100.
O
2
supply pressure at gas inlet is too
high.
O
2
supply pressure:
High
Check and connect gas supply lines.
Perform a Pre-use check.
O
2
supply pressure below 2.0 kPa x 100.
O
2
supply pressure at gas inlet is too low.
Gas supply line disconnected.
Note: This alarm can be permanently
silenced (Audio off) when activated.
O
2
supply pressure:
Low
Check control cable.
Contact a service technician.
No communication between user
interface and patient unit.
Panel disconnected
136
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
| 9 |
System messages
|
Summary of Contents for Maquet SERVO-s VENTILATOR SYSTEM V7.0
Page 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Page 2: ......
Page 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Page 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Page 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Page 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Page 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Page 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Page 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Page 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Page 183: ...x ...
Page 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...