Remedies
Possible causes
Alarm Message
Check patient breathing system.
Check patient connection (cuff
pressure/tracheal tube size).
Perform a Pre-use check.
Check ventilator settings.
Check alarm settings.
The measured end expiratory pressure
is above the preset or default alarm limit
for three consecutive breaths.
PEEP High
Check patient breathing system.
Check patient connection (cuff
pressure/tracheal tube size).
Perform a Pre-use check.
Check alarm settings.
The measured end expiratory pressure
is below the preset or default alarm limit
for three consecutive breaths.
Setting the alarm to zero turns the alarm
off.
Leakage in patient breathing system.
Leakage at patient connection (cuff,
tracheal tube).
PEEP Low
Check ventilator settings.
It is not possible to reach the Set volume
in PRVC due to restrictions imposed by
the set upper pressure limit.
Set high pressure alarm limit; this limits
the regulatory pressure used in PRVC.
Regulation pressure
limited
Attend to the patient.
Check the trigger setting.
Respiratory rate too high.
Auto triggering.
Respiratory Rate:
High
Attend to the patient.
Check trigger setting.
Check inspiratory cycle-off setting.
Respiratory rate too low.
Trigger sensitivity setting incorrect.
Large tidal volume.
Respiratory Rate:
Low
Check the adjustment for the inspiratory
tidal volume.
Setting causing larger volume than
allowed for the selected category.
Limited adjustment of excessive tidal
volume.
VT inspiratory
overrange
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
137
|
System messages
| 9 |
Summary of Contents for Maquet SERVO-s VENTILATOR SYSTEM V7.0
Page 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Page 2: ......
Page 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Page 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Page 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Page 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Page 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Page 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Page 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Page 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Page 183: ...x ...
Page 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...