11.2 VENTILATOR
11.2.1 GENERAL
Dimensions (mm)
Patient Unit: W 380 x D 300 x H 210
User Interface: W 355 x D 53 x H 295
- Screen type: TFT-LCD module
- Screen size: 31 cm (12.1") diagonal
- Viewing area: 246.0 x 184.5 mm
Weight, approximate (kg)
18
Triggering method
Flow and pressure
11.2.2 GAS SUPPLY
Gas quality
Supplied gases shall meet the requirements
for medical grade gases according to
applicable standards.
Maximum levels
Air
H
2
O < 7 g/m
3
Oil < 0.5 mg/m
3
Chlorine: must not be detectable
3
Oxygen
H
2
O < 20 mg/m
3
Oil < 0.3 mg/m
3
Inlet gas
Pressure: 2.0 – 6.0 kPa x 100 (29 – 87 psi)
max 60 l/min
Connection standards available
AGA, DISS, NIST, or French.
Unavailable gas/loss of gas pressure
The flow from an unavailable gas (air or O
2
) is
automatically compensated for so that the
patient gets the preset volume and pressure.
11.2.3 PATIENT SYSTEM CONNECTORS
Conical fittings (mm)
Male 22 and female 15, in accordance with
ISO 5356-1.
Gas exhaust port (mm)
Male 30 cone.
3. If the compressed air is generated by a liquid ring compressor there is a potential risk of chlorine in the supplied
air.
146
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
| 11 |
Technical data
|
Summary of Contents for Maquet SERVO-s VENTILATOR SYSTEM V7.0
Page 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Page 2: ......
Page 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Page 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Page 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Page 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Page 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Page 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Page 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Page 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Page 183: ...x ...
Page 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...