Setting range
Factory set default
Parameter
21 - 100
40
O
2
concentration (%)
0 - 50
5
PEEP (cmH
2
O)
2 - 20
5
PEEP in NIV (cmH
2
O)
(PEEP+1) - 50
15
Phigh (cmH
2
O)
-20 - 0
-
Press trig sensitivity level (cmH
2
O)
0 - (120 - PEEP)
20
Pressure level above PEEP (cmH
2
O)
0 - (32 - PEEP)
5
Pressure level above PEEP in NIV (cmH
2
O)
5- (120 -PEEP)
20
Pressure level above PEEP in backup (cmH
2
O)
5 - (32 – PEEP)
5
Pressure level above PEEP in NIV backup (cmH2O)
0 - (120 - PEEP)
0
PS above PEEP in Bivent (cmH
2
O)
0 - (120 - P
High
)
0
PS above Phigh in Bivent (cmH
2
O)
4 - 100
15
Resp rate in backup
1 - 60
5
SIMV frequency (b/min)
0.2 - 10
2
Thigh (s)
0.1 - 5
0.9
Ti (s)
0.1 - 5
0.9
Ti in backup (s)
100 - 2000
500
Tidal Volume (ml)
100 - 2000
500
Tidal volume in backup (ml)
0 - 30
10
Tpause (%)
0 - 1.5
0.4
Tpause (s)
0.2 - 10
2
TPEEP (s)
10 - 250
50
Weight (kg)
152
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
| 11 |
Technical data
|
Summary of Contents for Maquet SERVO-s VENTILATOR SYSTEM V7.0
Page 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Page 2: ......
Page 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Page 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Page 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Page 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Page 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Page 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Page 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Page 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Page 183: ...x ...
Page 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...