13.5 MENU KEY
Alarm profile setup
Alarm
By pressing Autoset in controlled
modes of ventilation the alarm limits
are automatically set for:
Pressure
Volume
Resp. Rate
PEEP
In spontaneous modes an alarm
setting for apnea time is available.
Note: Autoset is not possible in
NIV.
Alarm profile setup. Possible
selections:
Pressure (upper)
Minute volume (lower and upper)
Respiratory rate (lower and upper)
End expiratory pressure (lower and
upper)
Alarm sound level (10-100%)
Note: In NIV the alarm sound can
be permanently silenced (audio off).
Alarm:Profile
This shows alarms that have been activated. The list is in alphabetical order.
Alarm:History
Audio pause (Silence or pre-silence alarms)
Alarm:
Review
Review trends, event log or configuration.
Review:Trends
Review:Event log
Review:View
configuration
170
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
| 13 |
Appendix • User interface
|
Summary of Contents for Maquet SERVO-s VENTILATOR SYSTEM V7.0
Page 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Page 2: ......
Page 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Page 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Page 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Page 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Page 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Page 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Page 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Page 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Page 183: ...x ...
Page 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...