Direct Access Knobs - Safety
The four Direct Access Knob parameters are
displayed at the bottom of the screen with
color-coded bars that indicate whether the
parameter values are within
generally-recognized safety limits.
SVS-5003_XX
The figure above shows the following
components.
1.
A Direct Access Knob
2.
A yellow bar indicating the corresponding
parameter value is outside safety limits; a
text message is displayed accompanied
by an audible signal.
3.
A red bar indicating the corresponding
parameter value is significantly outside
safety limits; a text message is displayed
accompanied by an audible signal.
Note: When the defined safety limits for a
given parameter have been reached, the
Direct Access Knob becomes inoperative
for 2 seconds to indicate that a safety limit
has been reached.
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
25
|
System Overview
| 2 |
Summary of Contents for Maquet SERVO-s VENTILATOR SYSTEM V7.0
Page 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Page 2: ......
Page 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Page 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Page 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Page 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Page 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Page 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Page 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Page 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Page 183: ...x ...
Page 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...